[SM2123B]  Industrial grade online sewage pH sensor

SM2123B SM2123B

Technical Parameters

Technical parameterParameter value
BrandSONBEST
PH range4~14
PH accuracy±0.5
Applicable temperature range 0 to 65 ° C
Communication InterfaceRS485
Default baud rate9600 8 n 1
PowerDC6~24V 1A
Running temperature-30~85℃
Working humidity5%RH~90%RH

Product Size

SM2123B

Wiring instructions

Why choose this product?

SM2123B

Why choose this product?

SM2123B SM2123B

How to use?

SM2123B

Application solution

SM2123B

Product List

SM2123B

Pictures

SM2123B SM2123B

Precautions

1、电极在测量前必须用已知PH值的标准缓冲溶液进行定为校准,为取得更正确的结果,已知PH值要可靠,而且其PH值愈接近被测值愈好。
2、取下电极保护套后要注意,在塑料保护栅内的敏感玻璃泡不与硬物接触,任何破损和擦毛都会使电极失效。
3、测量完毕,不用时应将电极保护套套上,保护套内应放少量3.3mol/L氯化钾溶液,以保持电极球泡的湿润。
    4、电极的引出端,必须保持清洁和干燥,绝对防止输出两端短路,否则将导致测量结果失准或失效。
    5、电极避免长期浸在蒸馏水中货蛋白质溶液和酸性氟化物溶液中,并防止和有机硅油脂接触。
6、电极经长期使用后,如发现梯度略有极爱内地,则可把电极下端浸泡在4%HF(氢氟酸)中3-5秒钟,用蒸馏水洗净,然后在氯化钾溶液中浸泡,使之复新。
7、被测溶液中如含有易污染敏感球泡货堵塞液接界的物质,而使电极钝化,其现象是敏感梯度降低,货读数不准。如此,则应根据污染物质的性质,以适当溶液清洗,使之复新。
8、凡需测量能溶解聚碳酸树脂的有机溶剂,请选用PH符合电极中玻璃外壳型号电极。
    注:选用清洗剂时,如能溶解聚碳酸树脂的清洗液,如四氯化碳,三氯乙烯,四氢呋喃等,则可能把聚碳酸树脂溶解后,涂在敏感玻璃球泡上,而使电极失效,请慎用!

Communication Protocol

    The product uses RS485 MODBUS-RTU standard protocol format, all operation or reply commands are hexadecimal data. The default device address is 1 when the device leaves the factory, and the module or NON-Recorder default baud rate is 9600,8,n,1 ,but data recorder default baud rate is 115200 .
1. Read data (function code 0x03)
    Inquiry frame (hexadecimal), sending example: query 1 data of 1# device, the upper computer sends the command: 01 03 00 00 00 01 84 0A .
Address Function Code Start Address Data Length Check Code
010300 0000 0184 0A
    For the correct query frame, the device will respond with data: 01 03 02 00 79 79 A6 , response format:
Address Function Code LengthData 1Check Code
0103 0200 79 79 A6
    Data description: The data in the command is hexadecimal, take data 1 as an example, 00 79 is converted to decimal value as 121, assuming the data magnification is 100, then the real value is 121/100=1.21, Others and so on.
2. Common data address table
Configuration Address Register Address Register Description Data Type Value Range
4000100 00pHRead Only0~65535
4010100 64Model CodeRead/Write0~65535
4010200 65 total number of measuring points read/write 1~20
4010300 66device addressread/write1~249
4010400 67baud rateread/write0~6
4010500 68communication moderead/write1~4
4010600 69protocol typeread/write1~10


 Online shop 

 Contact Online 

Email: sale@sonbest.com

TYPE: SM2123B

WeChat: 13918126020

SKYPE: SOOBUU

Phone: 021-51083595

QQ1: 71182854 QQ2:613035997