[SY8206]  Post-type infrared temperature screening detector

SY8206

Technical Parameters

Technical parameterParameter value
BrandSONBEST
Temperature measurement range30~+45℃
Temperature measurement accuracy ±1 or %1 take the maximum value (maximum ±0.5℃)
Repeat measurement accuracy ±0.3 or %0.5 take the maximum value (maximum ±0.3℃)
Test object non-object
Optical Resolution 1.6:1
Spectral range 18-14um
Response time 100ms
Emissivity0.96
Communication InterfaceRS485
Default baud rate9600 8 n 1
PowerAC185~265V 1A
DisplayLED
Running temperature-30~85℃

Product Size

SY8206 SY8206 SY8206 SY8206

Wiring instructions

Why choose this product?

SY8206

How to use?

SY8206

Application solution

SY8206 SY8206

Product List

SY8206

Pictures

SY8206 SY8206 SY8206

Communication Protocol

    The product uses RS485 MODBUS-RTU standard protocol format, all operation or reply commands are hexadecimal data. The default device address is 1 when the device leaves the factory, and the module or NON-Recorder default baud rate is 9600,8,n,1 ,but data recorder default baud rate is 115200 .
1. Read data (function code 0x03)
    Inquiry frame (hexadecimal), sending example: query 1 data of 1# device, the upper computer sends the command: 01 03 00 00 00 01 84 0A .
Address Function Code Start Address Data Length Check Code
010300 0000 0184 0A
    For the correct query frame, the device will respond with data: 01 03 02 00 79 79 A6 , response format:
Address Function Code LengthData 1Check Code
0103 0200 79 79 A6
    Data description: The data in the command is hexadecimal, take data 1 as an example, 00 79 is converted to decimal value as 121, assuming the data magnification is 100, then the real value is 121/100=1.21, Others and so on.
2. Common data address table
Configuration Address Register Address Register Description Data Type Value Range
4000100 00temperatureRead Only0~65535
4010100 64Model CodeRead/Write0~65535
4010200 65 total number of measuring points read/write 1~20
4010300 66device addressread/write1~249
4010400 67baud rateread/write0~6
4010500 68communication moderead/write1~4
4010600 69protocol typeread/write1~10

software for test product
Setup Software
Development File

 Online shop 

 Contact Online 

Email: sale@sonbest.com

TYPE: SY8206

WeChat: 13918126020

SKYPE: SOOBUU

Phone: 021-51083595

QQ1: 71182854 QQ2:613035997